BSV (ZSV) for surgical lights

BSV (ZSV) for surgical lights (24 Volt)

Surgical lights according to DIN EN 60601-2-41 (VDE 0750 Part 2-41) and other essential lights to supplement the emergency power supply to 710.564.3 from an additional backup power supply with a switching time of up to 0.5 s can not continue automatically. NOPE Ruhstrat generally works without interruption, thus providing the highest level of safety in the operation.

For the changeover, the requirements apply to 710.537.6. Additional security source must be sized to supply at least three hours. It must be rated for only one hour if it is assured in conjunction with the security source, the minimum operating time of the surgical light of three hours total. The replacement system must ensure a operating time three hours through the connected battery system.

Provisions and obligations of this security power enshrined in the VDE 0100 part 710 - 11/2002.

Contact